Accéder au contenu
MilliporeSigma
Toutes les photos(1)

Key Documents

MPHL02GF2

Millipore

Durapore® 0.45 µm, Millipak® Capsule

Millipak® 20, sterile

Synonyme(s) :

Sterile Millipak-20 Filter Unit 0.45 μm 3/4 in. TC/TC

Se connecterpour consulter vos tarifs contractuels et ceux de votre entreprise/organisme


About This Item

Code UNSPSC :
23151806
eCl@ss :
32031690
Nomenclature NACRES :
NB.24

Matériaux

PVDF membrane
PVDF vent cap
polycarbonate
polycarbonate support

Niveau de qualité

Stérilité

irradiated
sterile

Gamme de produits

Millipak® 20

Caractéristiques

hydrophilic

Fabricant/nom de marque

Millipak®

Paramètres

0.35 bar max. differential pressure (5 psid) at 123 °C (Forward)
0.7 bar max. differential pressure (10 psid) at 25 °C (Reverse)
1.7 bar max. differential pressure (25 psid) at 80 °C (Forward)
10 L process volume
4.1 bar max. differential pressure (60 psid) at 25 °C (Forward)
5.2 bar max. inlet pressure (75 psi) at 25 °C
6.4 L/min flow rate at 2.45 bar (ΔP)

Technique(s)

bioburden reduction: suitable

Surface de filtration

100 cm2

Diam. du raccord d'entrée

3/4 in.

Diam. du raccord de sortie

3/4 in.

Impuretés

≤0.5 EU/mL (LAL test, Aqueous extraction)
<0.5  EU/mL USP bacterial endotoxins (LAL test, sample aqueous extraction)

Extractibles par gravimétrie

≤1 mg/device

Matrice

Durapore®

Dimension de pores

0.45 μm

Point de bulle

≥1790 mbar (26 psig), air with water at 23 °C

Raccords

inlet sanitary flange
outlet sanitary flange
(19 mm (3/4 in.) Sanitary Flange Inlet and Outlet)

Conditions d'expédition

ambient

Catégories apparentées

Description générale

Device Configuration: Gamma Gold Capsule
Number of stacked discs: 20 discs

Conditionnement

Blister

Autres remarques

These products are manufactured in a facility which adheres to FDA Good Manufacturing Practices.
Directions for Use

  • Organism Retention: Microorganism
  • Mode of Action: Filtration (size exclusion)
  • Application: BioProcessing
  • Intended Use: Reduction or removal of microorganism/bioburden
  • Instructions for Use: Please refer Wetting Instructions for Filter Units with Durapore® Membrane user guide
  • Storage Statement: Store in dry location
  • Disposal Statement: Dispose of in accordance with applicable federal, state and local regulations.

Notes préparatoires

Sterilization Method
3 autoclave cycles of 90 min @ 123 °C; not in-line steam sterilizable
This product was manufactured with a Durapore® membrane which meets the criteria for a "non-fiber releasing" filter as defined in 21 CFR 210.3 (b) (6).

Remarque sur l'analyse

Gravimetric Extractables: using sterilized device after a 200 mL flush and 24 hours in water at controlled room temperature
Will meet the USP Oxidizable Substances Test requirements after a water flush of ≥200 mL

Informations légales

Durapore is a registered trademark of Merck KGaA, Darmstadt, Germany
MILLIPAK is a registered trademark of Merck KGaA, Darmstadt, Germany

Vous ne trouvez pas le bon produit ?  

Essayez notre Outil de sélection de produits.


Certificats d'analyse (COA)

Recherchez un Certificats d'analyse (COA) en saisissant le numéro de lot du produit. Les numéros de lot figurent sur l'étiquette du produit après les mots "Lot" ou "Batch".

Déjà en possession de ce produit ?

Retrouvez la documentation relative aux produits que vous avez récemment achetés dans la Bibliothèque de documents.

Consulter la Bibliothèque de documents

Notre équipe de scientifiques dispose d'une expérience dans tous les secteurs de la recherche, notamment en sciences de la vie, science des matériaux, synthèse chimique, chromatographie, analyse et dans de nombreux autres domaines..

Contacter notre Service technique