Accéder au contenu
MilliporeSigma
Toutes les photos(1)

Key Documents

CMP471TS1

Millipore

Cartouche Milligard® II en PES de 1,2/0,45 µm

pore size 1.2/0.45 μm, cartridge nominal length 10 in. (25 cm)

Synonyme(s) :

Milligard PES 10" 1.2/0.45 μm Cartridge

Se connecterpour consulter vos tarifs contractuels et ceux de votre entreprise/organisme


About This Item

Code UNSPSC :
23151806
eCl@ss :
32031690

Matériaux

polyethersulfone membrane
polypropylene housing
polypropylene support
polypropylene vent cap

Niveau de qualité

Conformité réglementaire

meets FDA Indirect Food Additive requirements cited in 21 CFR 177-182 (all component materials)

Stérilité

non-sterile

Mode de stérilisation compatible

autoclavable compatible
steam-in-place compatible

Gamme de produits

EMPROVE® Filter
Milligard®

Caractéristiques

hydrophilic

Fabricant/nom de marque

Milligard®

Paramètres

1.38 bar max. differential pressure (20 psid) at 25 °C (Reverse)
5.52 bar max. differential pressure (80 psid) at 4-25 °C (Forward)

Technique(s)

prefiltration: suitable

Longueur

317.5 mm (12.5 in.)

Largeur

2.7 in.

Longueur nominale des cartouches

10 in. (25 cm)

Diamètre

69 mm (2.71 in.)

Surface de filtration

0.6 m2

Impuretés

<0.25 EU/mL bacterial endotoxins (LAL test, per device; aqueous extraction; meeting requirements of USP <85>, EP 2.6.14 and JP 4.01)

Matrice

Milligard® PES

pore size

1.2/0.45 μm pore size

Entrée

sample type liquid

Code de la cartouche

Code 7

Vous recherchez des produits similaires ? Visite Guide de comparaison des produits

Description générale

Device Configuration: Cartridge

Autres remarques

The product is manufactured in a facility which adheres to Good Manufacturing Practices. The manufacturing facility’s Quality Management System is approved by an accredited registering body to the appropriate ISO 9001 Quality Standard.
Directions for Use

  • Organism Retention: Microorganism
  • Mode of Action: Filtration (size exclusion)
  • Application: BioProcessing
  • Intended Use: Reduction or removal of microorganism/bioburden
  • Instructions for Use: Please refer wetting instructions of Filters with MILLIGARD® PES Membrane user guide
  • Storage Statement: Please refer Filters with MILLIGARD® PES Membrane user guide
  • Disposal Statement: Dispose of in accordance with applicable federal, state and local regulations.

Notes préparatoires

Sterilization Method
Autoclave wet 25x, 60 minute cycles at 128°C. In-line steam (wet): 10 forward cycles, 30 min.,
135 °C at ≤0.34 bar (5 psid) and 5 reverse cycles,
30 min., 135 °C at ≤0.07 bar (<1 psid)*.
*Devices were validated using steam at full differential pressure starting at 31 psid to a final differential pressure of 1 psid, resulting in a calculated average of 4 psid in forward and reverse directions.
This product was manufactured with materials that meet the criteria for a "non-fiber releasing" filter as defined in 21 CFR 210.3 (b) (6).

Informations légales

Emprove is a registered trademark of Merck KGaA, Darmstadt, Germany
MILLIGARD is a registered trademark of Merck KGaA, Darmstadt, Germany

Vous ne trouvez pas le bon produit ?  

Essayez notre Outil de sélection de produits.


Certificats d'analyse (COA)

Recherchez un Certificats d'analyse (COA) en saisissant le numéro de lot du produit. Les numéros de lot figurent sur l'étiquette du produit après les mots "Lot" ou "Batch".

Déjà en possession de ce produit ?

Retrouvez la documentation relative aux produits que vous avez récemment achetés dans la Bibliothèque de documents.

Consulter la Bibliothèque de documents

Notre équipe de scientifiques dispose d'une expérience dans tous les secteurs de la recherche, notamment en sciences de la vie, science des matériaux, synthèse chimique, chromatographie, analyse et dans de nombreux autres domaines..

Contacter notre Service technique