GC Analysis of Volatiles in White Wine on VOCOL® after SPME using a 75 μm Carboxen/PDMS Fiber

GC Analysis of Volatiles in White Wine on VOCOL® after SPME using a 75 μm Carboxen/PDMS Fiber suitable for GC, application for SPME

Materials

standard

Product No.
Description
Pricing

Ethyl hexanoate

analytical standard

CONDITIONS

sample/matrix

white wine + 25% NaCl

SPME fiber

Carboxen/PDMS, 75 μm (57318)

extraction

10 min, 40 °C (headspace)

desorption process

3 min at 290 °C

column

VOCOL, 30 m × 0.25 mm I.D., 1.5 μm (24205-U)

oven

40 °C (2 min) to 220 °C at 10 °C/min

inj. temp.

290 °C

detector

GC-MS, quadrupole, m/z = 31-240

carrier gas

helium, 35 cm/sec

injection

splitless/split, closed 1 min

liner

0.75 mm liner

sample

white wine

Description

Analysis Note

In this application, SPME was used for the analysis of volatiles in white wine using a headspace method. The presence of aroma compounds in white wine is one of the most important attributes of its final quality. A Carboxen/PDMS fiber was inserted into the headspace of the vial for 10 minutes at 40 °C and analyzed by GC/MS. The thick film of the VOCOL GC column helps focus the volatile analytes.