G005524

US EPA Method 524.2: GC Analysis of Volatiles on SPB-624 after Purge & Trap using "K" Trap

suitable for GC

Sign Into View Organizational & Contract Pricing


About This Item

technique(s)

gas chromatography (GC): suitable

test parameters

sample/matrix: each analyte at 10 ppb in 25 mL water
purge trap: VOCARB 3000 "K" (24940-U)
dry purge: 2 min
desorption preheat: 205 °C
desorption process: 40 mL/min at 210 °C for 2 min
bake: 260 °C for 10 min
transfer line/valve temp.: 110 °C
sample preparation: purge is 40 mL/min at 25 °C for 11 min
column: SPB-624, 30 m x 0.25 mm I.D., 1.4 μm (24255)
oven: 40 °C (2 min), 7 °C/min to 135 °C, 30 °C/min to 230 °C (3 min)
carrier gas: helium, 1.2 mL/min
injection: 30:1 split
liner: 0.75 mm I.D. SPME

application(s)

environmental
food and beverages

Analysis Note

Method 524.2 describes the analysis of volatile organic compounds in surface, ground, and drinking water. This application illustrates the analysis of the compounds listed in this method on a 30m x 0.25 mm I.D., 1.4 μm SPB-624 column using purge and trap / GC-MS.

Analyte

Description

    m-Xylene
    p-Xylene

Choose from one of the most recent versions:

Certificates of Analysis (COA)

Lot/Batch Number

Sorry, we don't have COAs for this product available online at this time.

If you need assistance, please contact Customer Support.

Already Own This Product?

Find documentation for the products that you have recently purchased in the Document Library.

Visit the Document Library

Our team of scientists has experience in all areas of research including Life Science, Material Science, Chemical Synthesis, Chromatography, Analytical and many others.

Contact Technical Service