HPLC Analysis of PAHs in Water on SUPELCOSIL™ LC-PAH after SPME using 100 μm PDMS Fiber
Matériaux
Fibre SPME
Réf. du produit
Description
Tarif
Fibres SPME Polydimethylsiloxane (PDMS)
df 100 μm(PDMS, for use with autosampler, needle size 24 gaCONDITIONS
sample/matrix
5 μL PAH mix (Cat. No. 4-8743) in 5 mL water
SPME fiber
polydimethylsiloxane, 100 μm (57301)
extraction
immersion, 30 min (rapid stirring)
desorption process
static, 200 μL acetonitrile:water, 40:60, 2 min
column
SUPELCOSIL LC-PAH, 15 cm × 4.6 mm I.D., 5 μm particles (58318)
gradient
0 min - 50% A; 5 min - 50% A; 30 min - 100% A; 45 min - 100% A
flow rate
0-2 min: 0.2 mL/min 2-45 min: 1.0 mL/min
mobile phase
[A] acetonitrile:[B] water
detector
UV, 254 nm
Description
Informations légales
SUPELCOSIL is a trademark of Sigma-Aldrich Co. LLC