HPLC Analysis of Azithromycin on Ascentis® Express C18 5μm
Matériaux
Colonne analytique
Réf. du produit
Description
Tarif
Composant de phase mobile
Réf. du produit
Description
Tarif
Étalon
Réf. du produit
Description
Tarif
CONDITIONS
column
Ascentis Express C18, 15 cm x 4.6 mm I.D., 5 μm particles (50537-U)
mobile phase
[A] 10 mM potassium phosphate, dibasic, pH 7.0 with phosphoric acid; [B] methanol; (20:80, A:B)
flow rate
1.5 mL/min
pressure
2558 psi (176 bar)
column temp.
45 °C
detector
UV, 210 nm
injection
10 μL
Description
Remarque sur l'analyse
A simple high-performance liquid chromatography (HPLC) method was developed for the determination of azithromycin in raw materials and pharmaceutical formulations without any pretreatment or derivatization step.
Informations légales
Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany