GC Analysis of Trichloroanisole (TCA) and Precursors in Red Wine on SLB®-5ms after SPME using 100 μm PDMS Fiber

Matériaux
Colonne analytique
Réf. du produit
Description
Tarif
Fibre SPME
Réf. du produit
Description
Tarif
Fibres SPME polydiméthylsiloxane (PDMS)
metal alloy fiber, df 100 μm(PDMS, needle size 23 ga, for use with autosamplerCONDITIONS
sample/matrix
1.5 mL red wine + 600 μL 5% potassium carbonate + 240 mg sodium chloride + 60 μL acetic anhydride
SPME fiber
metal fiber coated with 100 μm PDMS (57928-U)
extraction
headspace, 50 °C for 30 min., with stirring
desorption process
250 °C for 3 min
column
SLB-5ms, 30 m x 0.25 mm I.D., 0.25 μm (28471-U)
oven
50 °C (1 min), 25 °C/min to 280 °C
detector
ECD, 290 °C
carrier gas
helium, 1.5 mL/min, constant
liner
0.75 mm I.D. SPME
Description
Informations légales
SLB is a registered trademark of Merck KGaA, Darmstadt, Germany