Saltar al contenido
Merck
Todas las fotos(1)

Key Documents

G1007615

LC/MS Analysis of Nucleotides on SeQuant® ZIC-cHILIC

application for LC-MS

Iniciar sesiónpara Ver la Fijación de precios por contrato y de la organización


About This Item

technique(s)

LC/MS: suitable

test parameters

column: SeQuant® ZIC-cHILIC, 10 cm x 2.1 mm, 3.0 μm particles (1.50657)
mobile phase: [A] acetonitrile; [B] 50 mM ammonium acetate, pH 5.0 with acetic acid in water
gradient: 26 to 27% B in 10 min; to 27 to 35% B in 10 min; to 26% B in 0.1 min; hold at 26% B for 9.9 min
flow rate: 0.3 mL/min
pressure: 1029 psi (71 bar)
column temp.: 50 °C
detector: Q-TOF (ESI-negative), XIC of the following m/z values: AMP-346, ADP-426, ATP-506, CMP-322, CDP-402, (CTP-482, GMP-362, GDP-442, GTP-522, UMP-323, UDP-403, UTP-483)
injection: 10 μL
sample: 10-30 μg/mL of nucleotides (AMP, ADP, ATP, CMP, CDP, CTP, GMP, GDP, GTP, UMP, UDP, UTP) (80/20 acetonitrile/50 mM ammonium acetate in water, pH 5.3 with acetic acid)

suitability

application for LC-MS

application(s)

pharmaceutical (small molecule)

Analysis Note

Nucleotides are the building blocks of nucleic acids. They comprise a nitrogenous base, a sugar, and a phosphate group making them highly polar and ionic. This causes a challenge for typical reversed phase chromatography (RPC). These charged polar analytes are poorly retained by RPC, and thus typically chromatographed by either ion-exchange chromatography or RPC in conjunction with an ion-pairing reagent. Nucleotide analysis using ion-pair reagents or high concentration of phosphate buffers cause high background and ion source pollution for MS detection. HILIC separation is an alternative that permits sensitive MS detection and without the use of ion-pair reagents.

Legal Information

SeQuant is a registered trademark of Merck KGaA, Darmstadt, Germany

Analito

Descripción

    Adenosine 5′-monophosphate sodium salt, from yeast, ≥99%
    Adenosine 5′-diphosphate sodium salt, bacterial, ≥95% (HPLC)
    Adenosine 5′-triphosphate disodium salt hydrate, BioXtra, ≥99% (HPLC), from microbial
    Cytidine 5′-monophosphate, Sigma Grade, ≥99% (HPLC), synthetic, powder
    Cytidine 5′-diphosphate sodium salt hydrate, from yeast, ≥95%
    Cytidine 5′-triphosphate disodium salt, ~95%
    Guanosine 5′-monophosphate disodium salt hydrate, from yeast, ≥99%
    Guanosine 5′-diphosphate sodium salt, Type I, ≥96% (HPLC)
    Guanosine 5′-triphosphate sodium salt hydrate, ≥95% (HPLC), powder
    Uridine 5′-monophosphate, ≥98%
    Uridine 5′-(trihydrogen diphosphate) sodium salt from Saccharomyces cerevisiae, 95-100%
    Uridine 5′-triphosphate trisodium salt hydrate, from yeast, Type III, ≥96%
    Ver todo (12)

Elija entre una de las versiones más recientes:

Certificados de análisis (COA)

Lot/Batch Number

Lo sentimos, en este momento no disponemos de COAs para este producto en línea.

Si necesita más asistencia, póngase en contacto con Atención al cliente

¿Ya tiene este producto?

Encuentre la documentación para los productos que ha comprado recientemente en la Biblioteca de documentos.

Visite la Librería de documentos

Nuestro equipo de científicos tiene experiencia en todas las áreas de investigación: Ciencias de la vida, Ciencia de los materiales, Síntesis química, Cromatografía, Analítica y muchas otras.

Póngase en contacto con el Servicio técnico