G003595
GC Analysis of Agricultural Pesticides (Standard) on SLB®-5ms
suitable for GC
About This Item
technique(s)
gas chromatography (GC): suitable
test parameters
column: SLB-5ms, 30 m x 0.25 mm I.D, 0.25 μm (28471-U)
oven: 100 °C (1 min), 10 °C/min to 300 °C (5 min)
inj. temp.: 250 °C
MSD interface: 300 °C
scan range: selected ion monitoring (SIM), 7 monitoring groups used
carrier gas: helium, 1 mL/min constant
injection: 1.0 μL pulsed (20 psi until 0.20 min), splitless (1.0 min)
liner: 4 mm I.D., single taper
sample: 29-component pesticide mix plus 1 internal standard, each at 100 ppb in toluene
application(s)
agriculture
food and beverages
Analysis Note
standard solution was made in toluene to increase it stability and enhance response of some of the pesticides. The extra peaks in the first portion of the run are due to interferences from the solvent.