MTGR65010
Cápsula de filtración Aervent® 50
pore size 0.2 μm
Sinónimos:
EPA Est. 041237-FRA-001
About This Item
Productos recomendados
material
PTFE membrane
polypropylene
polypropylene housing
Quality Level
sterility
non-sterile
sterilization compatibility
gamma compatible
product line
Aervent®
feature
hydrophobic
manufacturer/tradename
Aervent®
parameter
2.1 bar max. differential pressure at 25 °C
4.1 bar max. inlet pressure (60 psi) at 25 °C
technique(s)
gas filtration: suitable
L
5.3 cm (2.1 in.)
W
2.4 in.
diam.
6.2 cm (2.4 in.)
filtration area
19.6 cm2
impurities
≤0.5 EU/mL bacterial endotoxins (LAL test, Aqueous extraction)
gravimetric extractables
≤1.0 mg
matrix
Aervent®
pore size
0.2 μm pore size
bubble point
≥1100 mbar (16 psig), nitrogen with 60/40% IPA/water at 23 °C
fitting
(3 mm (1/8 in.) NPT Inlet and Outlet)
General description
This Prostak™ microfiltration open-channel module features a Pore Size of 0.1 μm and a filtration area of 0.17 m2. PSVVAG021 is available in a 2 stack, making it applicable for bench-top, pilot, and production scale systems.
The modules can be integrity tested and steam sterilized for 20 cycles.
Prostak™ MF modules must be used with a support holder, which controls deflection and provides sanitary fittings for connection to external piping. Sanitary gaskets supplied with the module provide a leak-free connection between the module(s) and the holder.
Application
- Sterile venting of vessels and carboys
- Autoclave vacuum break lines
- Low volume sterile filtration of non-aqueous fluids
- Small fermenter inlet and exhaust gas filtration
Preparation Note
10 autoclave cycles of 30 min @ 130 °C; not in-line steam sterilizable
Analysis Note
Quantitative retention of 107 CFU/cm2Brevundimonas diminuta (ATCC® 19146) per ASTM F838-83 methodology
Other Notes
- Organism Retention: Microorganism
- Mode of Action: Filtration (size exclusion)
- Application: BioProcessing
- Intended Use: Reduction or removal of microorganism/bioburden
- Instructions for Use: Please refer Steam Sterilization & Integrity Testing Procedures guide and first page of Visual Inspection Guide
- Storage Statement: Store in dry location
- Disposal Statement: Dispose of in accordance with applicable federal, state and local regulations.
Legal Information
¿No encuentra el producto adecuado?
Pruebe nuestro Herramienta de selección de productos.
Certificados de análisis (COA)
Busque Certificados de análisis (COA) introduciendo el número de lote del producto. Los números de lote se encuentran en la etiqueta del producto después de las palabras «Lot» o «Batch»
¿Ya tiene este producto?
Encuentre la documentación para los productos que ha comprado recientemente en la Biblioteca de documentos.
Nuestro equipo de científicos tiene experiencia en todas las áreas de investigación: Ciencias de la vida, Ciencia de los materiales, Síntesis química, Cromatografía, Analítica y muchas otras.
Póngase en contacto con el Servicio técnico