Determination of Vitamin A and E Using HPLC with UV Detection Following the Chinese National Standard (GB) Method
Introduction
Following GB method for Vitamin A and Vitamin E as it is written with a 250 x 4.6 mm Ascentis® Express C30 column gives baseline resolution between the critical pair (gamma-, and beta-tocopherol; peaks 3 and 4). When a 150 x 4.6 mm Ascentis® Express C30 column is used, baseline resolution is still maintained with a separation that is complete in less than half of the time of the original method. Additionally, the flow rate can be doubled with the 250 x 4.6 mm Ascentis® Express C30 column, which gives a separation that is less than ten minutes compared to the original method. This application demonstrates how converting methods to superficially porous particle technology improves throughput.
![Determination of Vitamin A and E Using HPLC with UV Detection Following the Chinese National Standard (GB) Method Vitamin A or retinyl acetate](/deepweb/assets/sigmaaldrich/marketing/global/images/technical-documents/protocols/analytical-chemistry/small-molecule-hplc/vitamin-a/vitamin-a.jpg)
Vitamin A or retinyl acetate
![Determination of Vitamin A and E Using HPLC with UV Detection Following the Chinese National Standard (GB) Method General structure of tocopherol](/deepweb/assets/sigmaaldrich/marketing/global/images/technical-documents/protocols/analytical-chemistry/small-molecule-hplc/tocopherol-general-structure/tocopherol-general-structure.jpg)
General structure of tocopherol
![Determination of Vitamin A and E Using HPLC with UV Detection Following the Chinese National Standard (GB) Method Ascentis Express](/deepweb/assets/sigmaaldrich/marketing/global/images/technical-documents/protocols/analytical-chemistry/small-molecule-hplc/ascentis-express-c30-250cm-08ml-a/ascentis-express-c30-250cm-08ml-a.jpg)
![Determination of Vitamin A and E Using HPLC with UV Detection Following the Chinese National Standard (GB) Method Ascentis Express](/deepweb/assets/sigmaaldrich/marketing/global/images/technical-documents/protocols/analytical-chemistry/small-molecule-hplc/ascentis-express-c30-150cm-chromatogram/ascentis-express-c30-150cm-chromatogram.jpg)
![Determination of Vitamin A and E Using HPLC with UV Detection Following the Chinese National Standard (GB) Method Ascentis Express](/deepweb/assets/sigmaaldrich/marketing/global/images/technical-documents/protocols/analytical-chemistry/small-molecule-hplc/ascentis-express-c30-250cm-08ml/ascentis-express-c30-250cm-08ml.jpg)
![Determination of Vitamin A and E Using HPLC with UV Detection Following the Chinese National Standard (GB) Method Ascentis Express](/deepweb/assets/sigmaaldrich/marketing/global/images/technical-documents/protocols/analytical-chemistry/small-molecule-hplc/ascentis-express-c30-250cm-16ml/ascentis-express-c30-250cm-16ml.jpg)
Conclusion
Depending on the system capabilities, the separation throughput can be increased by using either a shorter length column or by doubling the flow rate.
Materials List
The column Ascentis® Express C30 250 x 4.6 mm, 2.7 µm is available as a custom product.