Skip to Content
Merck

US EPA Method 8141: GC Analysis of Organophosphorus Pesticides (OP-Pest) on Equity-5, NPD Detection

US EPA Method 8141: GC Analysis of Organophosphorus Pesticides (OP-Pest) on Equity-5, NPD Detection suitable for GC

Materials

analytical column

Product No.
Description
Pricing

Equity-5 Capillary GC Column

L × I.D. 30 m × 0.25 mm, df 0.25 μm

CONDITIONS

column

Equity-5, 30m × 0.25 mm I.D., 0.25 μm (28089-U)

oven

120 °C (3 min), 5 °C/min, to 300 °C (5 min)

inj. temp.

250 °C

detector

NPD, 320 °C

carrier gas

helium, 25 cm/sec @ 120 °C

injection

1.0 μL splitless (0.5 min)

liner

4 mm I.D., double taper

sample

40 ng on-column of a custom organophosphorus pesticides mix

Description

Legal Information

null