HPLC Analysis of Sugars and Taurine in an Energy Drink on Ascentis® Express HILIC using UV and ELSD Detection in Series

application for HPLC

Iniciar sesiónpara Ver la Fijación de precios por contrato y de la organización


About This Item

technique(s)

HPLC: suitable

test parameters

column: Ascentis Express HILlC, 10 cm x 3.0 mm I.D., 2.7 μm particles (53970-U)
mobile phase: [A] 100 mM ammonium acetate, pH 5.0 with acetic acid; [B] water; [C] acetonitrile; (09:01:90, A:B:C)
flow rate: 0.6 mL/min
pressure: 812 psi (56 bar)
column temp.: 35 °C
detector: ELSD, 55 °C, 3.5 bar nitrogen
injection: 2 μL
sample: dilute1:9 in acetonitrile

suitability

application for HPLC

application(s)

food and beverages

Analysis Note

This application demonstrates the suitability of Ascentis Express HILIC for the efficient separation of common drink additives used in caffeinated energy drinks. The analysis was performed with UV and ELSD detectors in series, this chromatogram being from the ELSD.

Legal Information

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany

Analito

Descripción

    Sucrose
    Taurine

Standard

Referencia del producto
Descripción
Precios

Elija entre una de las versiones más recientes:

Certificados de análisis (COA)

Lot/Batch Number

Lo sentimos, en este momento no disponemos de COAs para este producto en línea.

Si necesita más asistencia, póngase en contacto con Atención al cliente

¿Ya tiene este producto?

Encuentre la documentación para los productos que ha comprado recientemente en la Biblioteca de documentos.

Visite la Librería de documentos

Nuestro equipo de científicos tiene experiencia en todas las áreas de investigación: Ciencias de la vida, Ciencia de los materiales, Síntesis química, Cromatografía, Analítica y muchas otras.

Póngase en contacto con el Servicio técnico