Skip to Content
MilliporeSigma

HPLC Analysis of a Protein Mixture on Zenix® SEC-150 versus Zenix® SEC-300, Effect of Pore Size on Resolution

HPLC Analysis of a Protein Mixture on Zenix® SEC-150 versus Zenix® SEC-300, Effect of Pore Size on Resolution application for HPLC

Materials

Analytical column

Product No.
Description
Pricing

Zenix® SEC-150 Gel Filtration Column, 3 μm

L × I.D. 30 cm × 7.8 mm, 3 μm particle size, stainless steel

Zenix® SEC-300 Gel Filtration Column, 3 μm

L × I.D. 30 cm × 7.8 mm, 3 μm particle size, stainless steel

Standard

Product No.
Description
Pricing

γ-Globulins from human blood

≥99% (electrophoresis)

Thyroglobulin from bovine thyroid

powder, ≥90% (agarose gel electrophoresis)

Ribonuclease A from bovine pancreas

for molecular biology, ≥70 Kunitz units/mg protein, lyophilized

4-Aminobenzoic acid

analytical standard

Ovalbumin (257-264) chicken

≥97% (HPLC)

CONDITIONS

column

Zenix SEC-150, 30 cm x 7.8 mm I.D., 3 μm, 150 Å particles (Z777018)

column

Zenix SEC-300, 30 cm x 7.8 mm I.D., 3 μm, 300 Å particles (Z777033)

mobile phase

150 mM sodium phosphate, pH 7

flow rate

1 mL/min

column temp.

ambient

detector

UV, 214 nm

injection

10 μL

sample

protein mixture

Description

Legal Information

Zenix is a registered trademark of Sepax Technologies