跳轉至內容
Merck

LC/MS/MS Analysis of Pesticide Residues in Dates on the Ascentis® Express RP-Amide Column after QuEChERS Extraction

LC/MS/MS Analysis of Pesticide Residues in Dates on the Ascentis® Express RP-Amide Column after QuEChERS Extraction application for LC-MS, application for SPE

材料

分析柱

產品號碼
描述
訂價

Ascentis® Express RP-Amide, 2 μm UHPLC Column

2 μm particle size, L × I.D. 10 cm × 2.1 mm

基底

產品號碼
描述
訂價

草铵膦

PESTANAL®, analytical standard

標準

產品號碼
描述
訂價

喹恶磷

PESTANAL®, analytical standard

鱼藤酮

PESTANAL®, analytical standard

涕灭威砜

PESTANAL®, analytical standard

乙嘧硫磷

PESTANAL®, analytical standard

草不隆

PESTANAL®, analytical standard

氧乐果

PESTANAL®, analytical standard

丁酮威亚砜

PESTANAL®, analytical standard

环氧乙烷 溶液

certified reference material, 50 mg/mL in methanol

螺甲螨酯

PESTANAL®, analytical standard

马拉硫磷

certified reference material, TraceCERT®

多菌灵

certified reference material, TraceCERT®

涕灭威

certified reference material, TraceCERT®, Manufactured by: Sigma-Aldrich Production GmbH, Switzerland

噻虫啉

certified reference material, TraceCERT®, Manufactured by: Sigma-Aldrich Production GmbH, Switzerland

克百威

reference material

联苯肼酯

reference material

氯虫酰胺

reference material

噻唑隆

certified reference material, TraceCERT®, Manufactured by: Sigma-Aldrich Production GmbH, Switzerland

Spinetoram solution

100 μg/mL in acetonitrile, certified reference material, TraceCERT®, Manufactured by: Sigma-Aldrich Production GmbH, Switzerland

樣品製備

產品號碼
描述
訂價

Supel QuE

Citrate Extraction Tube, pk of 50, suitable for EN 15662

Supel QuE

Empty Centrifuge Tube with Lid, centrifuge tube volume 50 mL, pk of 50, suitable for EN 15662:2008 per BS

Supel QuE

PSA/C18 Tube, pack of 100, AOAC® 2007.01

Supel QuE

PSA Tube, pk of 50, AOAC® 2007.01, centrifuge tube volume 15 mL , Shaker Compatible

附屬品

產品號碼
描述
訂價

Certified Vial Kit, Low Adsorption (LA), 2 mL, pk of 100

volume 2 mL, amber glass vial (with marking spot), natural PTFE/silicone septa (with slit), thread for 9 mm

Benchmixer XL multi-tube vortexer

AC/DC input 115 V AC

CONDITIONS

sample/matrix

5 g ground dates spiked with pesticides at 10 ng/g

extraction process

add 10 mL water and 10 mL acetonitrile; shake 1 hr at 2500 rpm; add contents of Supel QuE citrate extraction tube (55227-U); . (shake for 1 minute; centrifuge at 5000 rpm fo 5 minutes and draw off supernatant)

clean-up tube

Supel QuE PSA/C18 (55288-U)

clean-up process

transfer 1 mL of the extract to 2 mLSupel QuE PSA/C18 tube; shake for 1 minute; centrifuge at 5000 rpm for 3 min; draw off supernatant for analysis.

sample preparation

Dilute 100 μL of cleaned extract to a final volume of 1 mL in 5 mM ammonium formate/0.1% formic acid in water; transfer diluted sample to vial for LC/MS/MS analysis

column

Ascentis Express RP-Amide, 10 cm x 2.1 mm I.D., 2 μm particles (51576-U)

mobile phase

(A) 5 mM ammonium formate with 0.1% formic acid in water; (B) 5 mM ammonium formate with 0.1% formic acid in 95:5 acetonitrile:H2O

gradient

held at 5% B from 0 to 1 min; to 100% B in 13 min; held at 100% B for 1.5 min; to 95% A in 0.5 min; held at 95% A for 1.5 min

flow rate

0.4 mL/min

column temp.

30 °C

detector

MS, ESI (+), MRM

injection

5 μL

描述

應用

Freeze dates with liquid nitrogen and grind to a homogenous mixture.

分析報告

The “quick, easy, cheap, effective, rugged, and safe” (QuEChERS) method can be used to analyze multiple pesticide residues found in dates. In this study, QuEChERS extraction and cleanup using PSA/C18 were used for the analysis of pesticide residues of interest in pistachios. Average recoveries obtained for replicate samples spiked at 10 ng/g were in the range of 70-120% for 23 of the 25 pesticides analyzed.

法律資訊

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany