跳轉至內容
Merck
  • Enzymatic determination of diglyceride using an iridium nano-particle based single use, disposable biosensor.

Enzymatic determination of diglyceride using an iridium nano-particle based single use, disposable biosensor.

Sensors (Basel, Switzerland) (2010-01-01)
Shu-Yi Hsu, Brandon Bartling, Christina Wang, Fuh-Sheng Shieu, Chung-Chiun Liu
摘要

A single use, disposable iridium-nano particle contained biosensor had been developed for the determination of diglyceride (DG). In this study hydrogen peroxide, formed through the enzymatic breakdown of DG via lipase, glycerol kinase and glycerol 3-phosphate oxidase, was electrochemically oxidized at an applied potential of +0.5 V versus the Ag/AgCl reference electrode. The oxidation current was then used to quantify the diglyceride concentration. Optimum enzyme concentrations and the surfactant loading used were established for successful sensor response. Good linear performance was observed over a DG concentration range of 0 to 25 μM in phosphate buffer and bovine serum media.

材料
產品編號
品牌
產品描述

Sigma-Aldrich
3-磷酸​​甘油氧化酶 来源于绿色气球菌, lyophilized powder, ≥70 units/mg solid
Sigma-Aldrich
3-磷酸​​甘油氧化酶 来源于小球菌 属, lyophilized powder, 40-80 units/mg solid, pH 8.1