G005854
LC/MS Analysis of Antiarrhythmic Drugs and Metabolites in Plasma on Ascentis® Express HILIC following Sample Prep using HybridSPE®-Phospholipid
application for HPLC
About This Item
technique(s)
HPLC: suitable
test parameters
sample preparation: SPE (Solid Phase Extraction)
sample/matrix: rabbit plasma, unfiltered K2-EDTA, spiked with each compound at 100 ng/mL (3:1, plasma:1% formic acid in acetonitrile)
SPE tube/cartridge: HybridSPE-Phospholipid, 96-well plate (575656-U)
sample addition: to each well add 100 μL plasma, followed by a 300 μL of 1% formic acid in acetonitrile
elution: collect, filtrate and analyze directly.
column: Ascentis Express HILIC, 10 cm x 2.1 mm I.D., 2.7 μm particles (53939-U)
mobile phase: [A] 5 mM ammonium formate; [B] 5 mM ammonium formate in acetonitrile, 5:95 (A:B), adjusted to pH 7.0 with formic acid
flow rate: 0.4 mL/min
pressure: 1305 psi (90 bar)
column temp.: 35 °C
detector: ESI+, 100-1000 m/z
injection: 0.5 μL
suitability
application for HPLC
application(s)
clinical
Analysis Note
Legal Information
Analyte
Choose from one of the most recent versions:
Certificates of Analysis (COA)
Sorry, we don't have COAs for this product available online at this time.
If you need assistance, please contact Customer Support.
Already Own This Product?
Find documentation for the products that you have recently purchased in the Document Library.
Our team of scientists has experience in all areas of research including Life Science, Material Science, Chemical Synthesis, Chromatography, Analytical and many others.
Contact Technical Service