Skip to Content
Merck

HPLC Analysis of Tricyclic Antidepressants in Serum on Discovery® C18 after SPE using Discovery® DSC-18

HPLC Analysis of Tricyclic Antidepressants in Serum on Discovery® C18 after SPE using Discovery® DSC-18 application for SPE, application for HPLC

Materials

analytical column

Product No.
Description
Pricing

Discovery® C18 HPLC Column

5 μm particle size, L × I.D. 15 cm × 4.6 mm

related product

Product No.
Description
Pricing

Discovery® DSC-18 SPE Tube

bed wt. 100 mg, volume 1 mL, pk of 108

CONDITIONS

sample preparation

SPE (Solid Phase Extraction)

SPE tube/cartridge

Discovery DSC-18, 100 mg/1 mL (52602-U)

column

Discovery C18, 15 cm × 4.6 mm, 5 μm preceded by a 2 cm C18 guard column and 0.5 μm frit filter (504955)

mobile phase

[A] acetonitrile: [B] methanol: [C] 25 mM potassium phosphate, monobasic (pH 7 with triethlylamine) (45:25:30, A:B:C)

flow rate

1 mL/min

column temp.

ambient

detector

UV, 254 nm

injection

50 μL, diluted porcine serum extract

Description

Legal Information

Discovery is a registered trademark of Merck KGaA, Darmstadt, Germany