Skip to Content
Merck

HPLC Analysis of Proteins on BIOshell A160 Peptide C18

HPLC Analysis of Proteins on BIOshell A160 Peptide C18 application for HPLC

Materials

analytical column

Product No.
Description
Pricing

Ascentis® Express Peptide ES-C18, 2.7 μm HPLC Column

2.7 μm particle size, L × I.D. 15 cm × 2.1 mm

standard

Product No.
Description
Pricing

Trypsin inhibitor from Glycine max (soybean)

lyophilized powder

Ribonuclease A from bovine pancreas

for molecular biology, ≥70 Kunitz units/mg protein, lyophilized

Lysozyme from chicken egg white

powder (crystalline), ~70000 U/mg

Superoxide Dismutase from bovine erythrocytes

lyophilized powder, ≥3,000 units/mg protein, Protein ≥95 % by biuret

Cytochrome c equine

recombinant, expressed in E. coli

CONDITIONS

column

BIOshell A160 Peptide C18, 15 cm x 2.1 mm I.D., 2.7 μm particles (66905-U)

mobile phase

[A] 75:25, 0.1% TFA in water: 0.08% TFA in acetonitrile; [B] 25:75, 0.1% TFA in water: 0.08% TFA in acetonitrile

gradient

held at 0% B for 0.5 min; 0 to 100% B in 7 min

flow rate

0.3 mL/min

pressure

4743 psi (327 bar)

column temp.

35 °C

detector

UV, 220 nm

injection

1 μL

sample

0.2 - 0.6 g/L in mobile phase A

Description

Legal Information

null