G005592
HPLC Analysis of Grapefruit Juice on Ascentis® Express C18 Column
application for HPLC, application for SPE
About This Item
technique(s)
HPLC: suitable
solid phase extraction (SPE): suitable
test parameters
sample preparation: SPE (Solid Phase Extraction)
sample/matrix: grapefruit juice hand-squeezed in laboratory, centrifuged and filtered through a 0.45 μm nylon filter disc (Acrodisc)
column: Ascentis Express C18, 5 cm x 4.6 mm I.D., 2.7 μm particles (53826-U)
mobile phase: [A] water/formic acid (99.9:0.1); [B] water/acetonitrile/isopropanol/formic acid (39.9:20:40:0.1)
gradient: 10% B for 3 min; 10 to 34% B in 30 min; 34 to 100% B in 1 min; held at 100% B for 6 min; 100 to10% B in 1 min
flow rate: 0.8 mL/min
column temp.: ambient
detector: PDA: 190–370 nm; chromatogram extracted at 283 nm
detector: ESI-MS: mass spectral range, 250-700 m/z; interval, 0.5 s; scan speed, 938 amu/s; nebulizing gas (N2) flow, 1.5 L/min; interface temperature, 350 °C; (heat block, 300 °C; desolvation line (DL) temperature, 300 °C; DL voltage, 34 V; probe voltage, +4.5 kV; Qarray voltage, 1.0 V; RF voltage, 90 V; detection gain, 1.05 kV.)
injection: 2 μL
suitability
application for HPLC
application for SPE
application(s)
food and beverages
Legal Information
Analyte
Analytical column
Standard
mobile phase component
Choose from one of the most recent versions:
Certificates of Analysis (COA)
Sorry, we don't have COAs for this product available online at this time.
If you need assistance, please contact Customer Support.
Already Own This Product?
Find documentation for the products that you have recently purchased in the Document Library.
Our team of scientists has experience in all areas of research including Life Science, Material Science, Chemical Synthesis, Chromatography, Analytical and many others.
Contact Technical Service