Przejdź do zawartości
Merck
Wszystkie zdjęcia(1)

Key Documents

CHGL02TP3

Millipore

Multilayer Durapore® 0.45/0.22 µm, Cartridge

cartridge nominal length 20 in. (50 cm), filter diam. 2.7 in. (6.9 cm), Code 0

Zaloguj sięWyświetlanie cen organizacyjnych i kontraktowych


About This Item

Kod UNSPSC:
23151806

Materiały

polypropylene device
polypropylene support
silicone seal

Poziom jakości

zgodność regionalna

meets FDA 21 CFR 177-182 (all component materials)

sterylność

non-sterile

Kompatybilność ze sterylizacją

autoclavable compatible
steam-in-place compatible

linia produktu

Durapore®

Właściwości

hydrophilic

producent / nazwa handlowa

Durapore®

Parametry

≤21.6 mL/min air diffusion at 2.75 bar (40 psig) (in water)
0.35 bar max. differential pressure (5 psid) at 135 °C (forward)
0.35 bar max. inlet pressure (5 psi) at 135 °C (forward)
1.7 bar max. differential pressure (25 psid) at 80 °C (forward)
1.7 bar max. inlet pressure (25 psi) at 80 °C (forward)
3.5 bar max. differential pressure (50 psid) at 25 °C (reverse; intermittent)
3.5 bar max. inlet temp. (50 psi) at 25 °C (reverse; intermittent)
5.5 bar max. differential pressure (80 psid) at 25 °C (forward)
5.5 bar max. inlet pressure (80 psi) at 25 °C (forward)

metody

sterile filtration: suitable

Długość

20 in.

Szerokość

2.7 in.

długość nominalna kartridża

20 in. (50 cm)

device size

20 in.

śr. filtra

2.7 in. (6.9 cm)

filter filtration area

1.1 m2

zanieczyszczenia

≤0.25 EU/mL bacterial endotoxins (LAL test, aqueous extraction)
<0.25 EU/mL USP bacterial endotoxins (LAL test, sample aqueous extraction)

macierz

Multilayer Durapore®

pore size

0.45/0.22 μm pore size

punkt pęcherzykowania

≥3450 mbar (50 psig), air with water

kod

Code 0

Szukasz podobnych produktów? Odwiedź Przewodnik dotyczący porównywania produktów

Opis ogólny

Device configuration: Cartridge
Cartridge type: Liquid

Uwaga dotycząca przygotowania

Sterilization Method: 30 SIP cycles of 30 min at 135°C; 30 autoclave cycles of 60 min at 126°C

This product was manufactured with a Durapore®   membrane which meets the criteria for a "non-fiber releasing" filter as defined in 21 CFR 210.3 (b) (6).

Komentarz do analizy

Bacterial Retention: Samples were quantitatively retentive of a minimum Brevundimonas diminuta challenge concentration of 1 x 107 CFU/cm2  using ASTM® F838 methodology.
Will meet the USP Oxidizable Substances Test requirements after a water flush of ≥ 3L

Inne uwagi

Directions for Use:
  • Organism Retention: Microorganism
  • Mode of Action: Filtration (size exclusion)
  • Application: BioProcessing
  • Intended use: Reduction or removal of microorganism/bioburden
  • Instructions for Use: Please refer to Wetting Instructions for Filter Units with Durapore® Membrane user guide
  • Storage Statement: Store in dry location
  • Disposal Statement: Dispose of in accordance with applicable federal, state and local regulations

Informacje prawne

ASTM is a registered trademark of American Society for Testing and Materials
Durapore is a registered trademark of Merck KGaA, Darmstadt, Germany
This page may contain text that has been machine translated.

Nie możesz znaleźć właściwego produktu?  

Wypróbuj nasz Narzędzie selektora produktów.


Certyfikaty analizy (CoA)

Poszukaj Certyfikaty analizy (CoA), wpisując numer partii/serii produktów. Numery serii i partii można znaleźć na etykiecie produktu po słowach „seria” lub „partia”.