Skip to Content
Merck

HPLC Analysis of PAHs in Water on SUPELCOSIL LC-PAH after SPME using 100 μm PDMS Fiber

HPLC Analysis of PAHs in Water on SUPELCOSIL™ LC-PAH after SPME using 100 μm PDMS Fiber application for SPME, application for HPLC

Materials

SPME fiber

Product No.
Description
Pricing

SPME fiber assembly Polydimethylsiloxane (PDMS)

df 100 μm(PDMS, for use with autosampler, needle size 24 ga

CONDITIONS

sample/matrix

5 μL PAH mix (Cat. No. 4-8743) in 5 mL water

SPME fiber

polydimethylsiloxane, 100 μm (57301)

extraction

immersion, 30 min (rapid stirring)

desorption process

static, 200 μL acetonitrile:water, 40:60, 2 min

column

SUPELCOSIL LC-PAH, 15 cm × 4.6 mm I.D., 5 μm particles (58318)

gradient

0 min - 50% A; 5 min - 50% A; 30 min - 100% A; 45 min - 100% A

flow rate

0-2 min: 0.2 mL/min 2-45 min: 1.0 mL/min

mobile phase

[A] acetonitrile:[B] water

detector

UV, 254 nm

Description

Legal Information

SUPELCOSIL is a trademark of Sigma-Aldrich Co. LLC