Saltar al contenido
Merck
HomeProtein PurificationConverting from Flow Velocity to Volumetric Flow Rates

Converting from Flow Velocity to Volumetric Flow Rates

It is convenient when comparing results for columns of different sizes to express flow as flow velocity (cm/h). However, flow is usually measured in volumetric flow rate (ml/min). To convert between flow velocity and volumetric flow rate use one of the formulae below.

From flow velocity (cm/h) to volumetric flow rate (ml/min)

Volumetric flow rate (ml/min) = Flow velocity (cm/h)
60
× column cross sectional area (cm2)

Volumetric flow rate (ml/min) = Y
60
× π × d2
4

where

Y = flow velocity in cm/h
d = column inner diameter in cm

Example:

What is the volumetric flow rate in an XK 16/70 column (i.d. 1.6 cm) when the flow velocity is 150 cm/h?
Y = flow velocity = 150 cm/h
d = inner diameter of the column = 1.6 cm

Volumetric flow rate (ml/min) = 150 × π × 1.6 × 1.6
60 × 4
= 5.03 ml/min

From volumetric flow rate (ml/min) to flow velocity (cm/h)

flow velocity (cm/h) = Volumetric flow rate (ml/min) × 60
column cross sectional area (cm2)

flow velocity (cm/h) = = Z × 60 × 4
π × d2

where
Z = volumetric flow rate in ml/min
d = column inner diameter in cm

Example:

What is the linear flow in a Tricorn 5/50 column (i.d. 0.5 cm) when the volumetric flow rate is 1 ml/min?
Z = volumetric flow rate = 1 ml/min
d = column inner diameter = 0.5 cm

flow velocity = 1 × 60 × 4
π × 0.5 × 0.5
= 305.6 cm/h

From volumetric flow rate (ml/min) to using a syringe

1 ml/min = approximately 30 drops/min on a HiTrap 1 ml column
5 ml/min = approximately 120 drops/min on a HiTrap 5 ml column

Materials
Loading
Inicie sesión para continuar.

Para seguir leyendo, inicie sesión o cree una cuenta.

¿No tiene una cuenta?