Saltar al contenido
Merck

GC Analysis of Resveratrol in Red Wine on SLB®-5ms after SPME using 85 μm Polyacrylate Fiber

GC Analysis of Resveratrol in Red Wine on SLB®-5ms after SPME using 85 μm Polyacrylate Fiber suitable for GC, application for SPME

Materiales

Related product

Referencia del producto
Descripción
Precios

Portafibra SPME

for use with manual sampling

SPME/Direct

pkg of 1 ea

SPME fiber

Referencia del producto
Descripción
Precios

SPME fiber assembly polyacrylate (PA)

df 85 μm(PA, for use with manual holder, needle size 24 ga

SPME fiber assembly polyacrylate (PA)

df 85 μm(PA, for use with autosampler, needle size 24 ga

SPME fiber assembly polyacrylate (PA)

df 85 μm(PA, for use with autosampler, needle size 23 ga

Standard

Referencia del producto
Descripción
Precios

Resveratrol

≥99% (HPLC)

CONDITIONS

sample/matrix

3 mL of red wine (California merlot) diluted 3:1 in 12% ethanol:water

SPME fiber

85 μm polyacrylate FS 23 gauge auto (57294-U)

SPME fiber

85 μm polyacrylate FS 24gauge auto (57305)

SPME fiber

85 μm polyacrylate FS 24gauge manual (57304)

extraction

immersion for 15 min at room temperature while stirring at 400 rpm, fiber gently blotted with a Kimwipe to remove excess water, (then derivatization by immersion in a 4 mL vial containing 5 μL of Sylon-BFT (fresh vial for each analysis than was allowed to equilabrate 60-90 minutes prior to use) for 20 min)

desorption process

280 °C for 2 min

column

SLB-5ms; 30 m x 0.25 mm I.D., 0.25 μm (28471-U)

oven

100 °C (1 min), 10 °C/min to 325 °C (3 min)

MSD interface

325 °C

scan range

m/z 40-450

carrier gas

helium, 1 mL/min, constant

liner

0.75 mm I.D., SPME

Descripción

Legal Information

SLB is a registered trademark of Merck KGaA, Darmstadt, Germany