Saltar al contenido
Merck

GC Analysis of Aromatics in Reformulated Gasoline on SUPELCOWAX 10

GC Analysis of Aromatics in Reformulated Gasoline on SUPELCOWAX 10 suitable for GC

Materiales

Analytical column

Referencia del producto
Descripción
Precios

Columna para GC capilar SUPELCOWAX 10

L × I.D. 60 m × 0.25 mm, df 0.25 μm

CONDITIONS

column

SUPELCOWAX 10, 60 m x 0.25 mm I.D., 0.25 μm (24081)

oven

50 °C (3 min), 15 °C/min to 265 °C (5 min)

inj. temp.

250 °C

detector

FID, 265 °C

carrier gas

helium, 30 cm/sec

injection

0.5 μL, 200:1 split

liner

4 mm I.D., splitless type, single taper wool packed FocusLiner design

sample

reformulated unleaded gasoline

Descripción

Analysis Note

This application illustrates the analysis of a sample of reformulated gasoline on the SUPELCOWAX 10 column. Due to the polar nature of the phase, the aliphatics elute earlier than the aromatics. There is some overlap in the elution ranges, however it is not as severe as that exhibited by a nonpolar column.

Legal Information

FocusLiner is a trademark of SGE Analytical Science Pty Ltd.