796-0150
HPLC Analysis of Acidic Herbicides in Water on a Polymeric C18 Column after SPE using Supelclean™ ENVI™-Carb
application for SPE, application for HPLC
About This Item
Productos recomendados
technique(s)
HPLC: suitable
solid phase extraction (SPE): suitable
test parameters
sample preparation: SPE (Solid Phase Extraction)
sample/matrix: Fresh 1 L water samples, dechlorinated with sodium thiosulfate when necessary, at ambient temperature and pH.
SPE tube/cartridge: Supelclean ENVI-Carb, 250 mg/6 mL (57092)
condition: 10 mL DI water at 20 mL/min
sample addition: 0.9 L water sample at 20 mL/min
drying: 10 min using clean nitrogen
washing: 10 mL DI water at 20 mL/min
elution: 10 mL 0.1% phosphoric acid in methylene chloride:acetonitrile (80:20) at 5 mL/min
column: polymeric-coated silica-based PAH specialty column, 20 cm × 3 mm I.D., 5 μm (Supelco equivalent, SUPELCOSIL LC-PAH, available upon request)
mobile phase: gradient, [A]: 0.05% phosphoric acid in DI water; [B]: acetonitrile
flow rate: 0.5 mL/min
column temp.: 50 °C
detector: photodiode array- peak width: 0.053 min, sampling interval: 0.320 sec, monitor 210 nm & 225 nm
injection: 10 μL of extract (4-5 ppb each analyte in water)
suitability
application for HPLC
application for SPE
application(s)
environmental
Application
Legal Information
Analito
SPE tube or plate
analytical column
Elija entre una de las versiones más recientes:
Certificados de análisis (COA)
Lo sentimos, en este momento no disponemos de COAs para este producto en línea.
Si necesita más asistencia, póngase en contacto con Atención al cliente
¿Ya tiene este producto?
Encuentre la documentación para los productos que ha comprado recientemente en la Biblioteca de documentos.
Nuestro equipo de científicos tiene experiencia en todas las áreas de investigación: Ciencias de la vida, Ciencia de los materiales, Síntesis química, Cromatografía, Analítica y muchas otras.
Póngase en contacto con el Servicio técnico