HPLC Analysis of Citalopram Enantiomers on Astec® CHIROBIOTIC® V
![HPLC Analysis of Citalopram Enantiomers on Astec® CHIROBIOTIC® V application for HPLC](/deepweb/assets/sigmaaldrich/product/images/238/810/37aad6d3-7cd5-4c00-b6a6-f5bfa0e11f7b/800/37aad6d3-7cd5-4c00-b6a6-f5bfa0e11f7b.jpg)
CONDITIONS
column
CHIROBIOTIC V, 25 cm x 4.6 mm I.D., 5 μm particles (11024AST)
mobile phase
[A] methanol; [B] acetic acid; [C] triethylamine; (99.8:0.1:0.1, A:B:C)
flow rate
0.5 mL/min
column temp.
20 °C
detector
UV, 213 nm
injection
5 μL
sample
citalopram, 1 mg/mL in mobile phase