Skip to Content
Merck

HPLC Analysis of Proteins on Proteomix® SCX-NP1.7, Demonstration of Lot-to-Lot Reproducibility

HPLC Analysis of Proteins on Proteomix® SCX-NP1.7, Demonstration of Lot-to-Lot Reproducibility application for HPLC

Materials

standard

Product No.
Description
Pricing

Lysozyme from chicken egg white

lyophilized powder, protein ≥90 %, ≥40,000 units/mg protein

Cytochrome c from equine heart

≥95% (SDS-PAGE)

Ribonuclease A from bovine pancreas

for molecular biology, ≥70 Kunitz units/mg protein, lyophilized

CONDITIONS

column

Proteomix SCX-NP1.7, 5 cm x 4.6 mm I.D., 1.7 μm particles (Z777146)

mobile phase

(A) 20 mM sodium phosphate buffer, pH 6.0, (B) A + 1.0 M NaCl

gradient

0 - 75% B in 25 min, 15 min prewash

flow rate

0.75 mL/min

column temp.

ambient

detector

UV, 214 nm

injection

5 μL

sample

1 mg/mL each compound

Description

Legal Information

Proteomix is a registered trademark of Sepax Technologies