Affinity Chromatography Troubleshooting
This section focuses on practical problems that may occur when running a chromatography column. The diagrams below give an indication of how a chromatogram may deviate from the ideal during affnity purifcation and what measures can be taken to improve the results.
Target elutes as a sharp peak. Satisfactory result
- If it is diffcult or impossible to retain biological activity when achieving this result, either new elution conditions or a new ligand must be found.
- If using low pH for elution, collect the fractions in neutralization buffer (60–200 µL 1 M Tris-HCl, pH 9.0 per mL eluted fraction).
Target is a broad, low peak that elutes while binding buffer is being applied
- Find better binding conditions.
Target elutes in a broad, low peak
- Try different elution conditions.
- If using competitive elution, increase the concentration of the competitor in the elution buffer.
- Stop flow intermittently during elution to allow time for the target molecule to elute and so collect the target protein in pulses (see second figure beneath).
Note: This result may also be seen if the target protein has denatured and aggregated on the column or if there is non-specifc binding.
Some of the target molecule elutes as a broad, low peak while still under binding conditions
- Allow time for the sample to bind and/or apply sample in aliquots, stopping the flow for a few minutes between each sample application (see second fgure beneath).
Materials
Loading
계속하려면 로그인하세요.
계속 읽으시려면 로그인하거나 계정을 생성하세요.
계정이 없으십니까?