Conditions

desorption process260 °C for 3 min.
extractionheadspace, 65 °C (30 min)
SPME fiber2 cm Metal coated with 50/30 µm DVB/Carboxen/PDMS (57914-U)
columnSLB-5ms, 30 m x 0.25 mm I.D., 0.25 µm (28471-U)
oven60 °C (2 min), 8 °C/min to 200 °C
carrier gashelium, 1 mL/min constant
sample20/10 ppt odor compounds in 25 mL water + 25% NaCl
liner0.75 mm I.D., SPME
MSD interface300 °C
scan rangeSIM, m/z = 137, 124, 95, 112