US EPA Method 8015 (modified): GC Analysis of Gasoline Range Organics (GRO) on Equity-1 after Purge & Trap using "M" Trap

US EPA Method 8015 (modified): GC Analysis of Gasoline Range Organics (GRO) on Equity-1 after Purge & Trap using "M" Trap suitable for GC

Materials

๋ถ„์„ ์ปฌ๋Ÿผ

์ œํ’ˆ ๋ฒˆํ˜ธ
์„ค๋ช…
๊ฐ€๊ฒฉ

SLBยฎ-5ms Capillary GC Column

L ร— I.D. 20 m ร— 0.18 mm, df 0.18 ฮผm

CONDITIONS

sample/matrix

Each analyte at 50 ppb in 5 mL of water

purge trap

MOD BTEXTRAP M, Tekmarยฎ 3000

purge

40 mL/min at 25 ยฐC for 11 min

dry purge

2 min

desorption preheat

220 ยฐC

bake

260 ยฐC for 10 min

transfer line/valve temp.

110 ยฐC

column

Equity-1, 30 m x 0.32 mm I.D., 1.0 μm (28057-U)

oven

40 ยฐC (2 min), 5 ยฐC/min to 75 ยฐC, 10 ยฐC/min to 200 ยฐC

inj. temp.

200 ยฐC

detector

FID, 250 ยฐC

carrier gas

helium, 1.4 mL/min constant

injection

20:1 split

liner

0.75 mm I.D. SPME

์„ค๋ช…

๋ฒ•์  ์ •๋ณด

null
Tekmar is a registered trademark of Tekmar Co.