콘텐츠로 건너뛰기
Merck

HPLC Analysis of Triacylglycerols in Olive Oil on Discovery® HS C18 after Sample Prep using Discovery® DSC-Si (Propionitrile Mobile Phase, ELSD Detection)

HPLC Analysis of Triacylglycerols in Olive Oil on Discovery® HS C18 after Sample Prep using Discovery® DSC-Si (Propionitrile Mobile Phase, ELSD Detection) application for HPLC

CONDITIONS

SPE tube/cartridge

Discovery DSC-Si, 1 gm/6 mL (52656-U)

condition

6 mL hexane

sample addition

0.12 g oil in 0.5 mL hexane

elution

10 mL hexane-diethyl ether 87-13 (v/v)

eluate post-treatment

evaporation to dryness, reconstituted into 2 mL acetone

column

Discovery HS C18, 25 cm x 4.6 mm I.D., 5 μm particles (568523-U)

mobile phase

propionitrile

flow rate

1.5 mL/min

pressure

1117 psi (77 bar)

column temp.

30 °C

detector

ELSD

injection

5 μL

설명

분석 메모

Olive oil was pre-cleaned using Silica SPE, analyzed as 5% sample in acetone (w/v).

법적 정보

Discovery is a registered trademark of Merck KGaA, Darmstadt, Germany