HPLC Analysis of Sugars on Ascentis® Si
![HPLC Analysis of Sugars on Ascentis® Si application for HPLC](/deepweb/assets/sigmaaldrich/product/images/246/049/1366241a-9ed0-4084-a1b1-051450f73a70/800/1366241a-9ed0-4084-a1b1-051450f73a70.jpg)
Materials
분석 컬럼
제품 번호
설명
가격
표준 물질
제품 번호
설명
가격
CONDITIONS
column
Ascentis Si, 15 cm x 4.6 mm I.D., 5 μm particles (581512-U)
mobile phase
[A] water; [B] acetonitrile; (15:85, A:B)
flow rate
1.0 mL/min
column temp.
35 °C
detector
ELSD, 55 °C, 3.5 bar nitrogen
injection
10 μL
설명
분석 메모
This application demonstrates the suitability of the Ascentis Si for the analysis of sugars. Structures along with the optimized chromatogram obtained on the Ascentis Si are presented. Separation is by normal-phase mechanism as a decrease in mobile phase water content results in increased retention.