US EPA Method 8082: GC Analysis of Aroclor 1016/1260 (PCBs) on SLB®-5ms

Materials
가아드 컬럼
부속물
제품 번호
설명
가격
Wheaton wide-mouth bottle
amber soda-lime glass bottle, capacity 1000 mL, white polypropylene cap, (PTFE faced LDPE liner), case of 12 ea분석 컬럼
제품 번호
설명
가격
추천
제품 번호
설명
가격
표준 물질
제품 번호
설명
가격
Pesticide Surrogate Spike Mix
certified reference material, TraceCERT®, 200 μg/mL each component in acetone, ampule of 1 mLCONDITIONS
column
SLB-5ms, 30 m x 0.25 mm I.D., 0.25 μm (28471-U)
oven
100 °C (2 min), 15 °C/min to 330 °C (3 min)
inj. temp.
250 °C
detector
ECD, 330 °C
carrier gas
helium, 25 cm/sec, constant
injection
2 μL splitless (0.75 min)
liner
4 mm I.D., single taper
sample
Aroclor standard mix 1 (46846-U) diluted to 500 ppb / 50 ppb (Aroclors / surrogates) in n-hexane
설명
분석 메모
Aroclors are commercial mixtures of PCB congeners. Due to their stability, they ave been used in many different industrial and commercial applications. Because of their toxicity and ability to bioaccumulate, production of these materials ceased in 1977. This application demonstrates the separation of two common Aroclor mixtures, Aroclor 1016 and Aroclor 1260 on the SLB-5ms.
법적 정보
SLB is a registered trademark of Merck KGaA, Darmstadt, Germany