콘텐츠로 건너뛰기
Merck

GC Analysis of Nitrogen-Containing Herbicides in Water on a 5% Phenyl Column after SPME using 85 μm Polyacrylate Fiber

GC Analysis of Nitrogen-Containing Herbicides in Water on a 5% Phenyl Column after SPME using 85 μm Polyacrylate Fiber suitable for GC, application for SPME

Materials

SPME 파이버

제품 번호
설명
가격

SPME fiber assembly polyacrylate (PA)

df 85 μm(PA, for use with manual holder, needle size 24 ga

CONDITIONS

sample/matrix

4 mL (water + 100ng/mL each analyte 1g/mL NaCl), pH 2, in 4.6 mL vial

SPME fiber

polyacrylate, 85 μm (57304)

extraction

immersion, ambient temp., 50 min (constant stirring)

desorption process

5 min, 230 °C

column

5% phenyl polysiloxane, 30 m × 0.25 mm I.D., 0.25 μm (Supelco equivalent, SLB-5ms, 28471-U)

oven

40 °C (5 min) to 100 °C at 30 °C/min, to 275 °C at 5 °C/min

inj. temp.

splitless, 230 °C

detector

MS (Ion Trap Manifold: 250°C; scan range m/z=45-400 at 0.6 sec/scan)

carrier gas

helium, 40 cm/sec, set at 40 °C