콘텐츠로 건너뛰기
Merck
모든 사진(1)

Key Documents

796-0086

HPLC Analysis of PAHs in Water on SUPELCOSIL LC-PAH after SPME using 100 μm PDMS Fiber

application for SPME, application for HPLC

로그인조직 및 계약 가격 보기


About This Item

기술

HPLC: suitable
solid phase microextraction (SPME): suitable

검사 파라미터

sample/matrix: 5 μL PAH mix (Cat. No. 4-8743) in 5 mL water
SPME fiber: polydimethylsiloxane, 100 μm (57301)
extraction: immersion, 30 min (rapid stirring)
desorption process: static, 200 μL acetonitrile:water, 40:60, 2 min
column: SUPELCOSIL LC-PAH, 15 cm × 4.6 mm I.D., 5 μm particles (58318)
gradient: 0 min - 50% A; 5 min - 50% A; 30 min - 100% A; 45 min - 100% A
flow rate: 0-2 min: 0.2 mL/min 2-45 min: 1.0 mL/min
mobile phase: [A] acetonitrile:[B] water
detector: UV, 254 nm

적합성

application for HPLC
application for SPME

응용 분야

environmental

법적 정보

SUPELCOSIL is a trademark of Sigma-Aldrich Co. LLC

분석물

설명

    Chrysene
    Naphthalene
    Benzo[b]fluoranthene
    Phenanthrene
    Indeno[1,2,3-cd]pyrene
    Benzo[a]pyrene
    Fluoranthene
    Anthracene
    Dibenz[a,h]anthracene
    Pyrene
    Acenaphthylene
    Benzo[k]fluoranthene
    Acenaphthene
    Fluorene
    Benz[a]anthracene
    모두 보기 (15)

가장 최신 버전 중 하나를 선택하세요:

시험 성적서(COA)

Lot/Batch Number

죄송합니다. 지금은 이 제품에 대한 COA이(가) 온라인에서 제공되지 않습니다.

도움이 필요하시면 연락하세요. 고객 지원 부서

이 제품을 이미 가지고 계십니까?

문서 라이브러리에서 최근에 구매한 제품에 대한 문서를 찾아보세요.

문서 라이브러리 방문

자사의 과학자팀은 생명 과학, 재료 과학, 화학 합성, 크로마토그래피, 분석 및 기타 많은 영역을 포함한 모든 과학 분야에 경험이 있습니다..

고객지원팀으로 연락바랍니다.