Skip to Content
Merck

HPLC Analysis of Cytochrome c Species Variants on Discovery® BIO PolyMA-SCX (Ion Exchange)

HPLC Analysis of Cytochrome c Species Variants on Discovery® BIO PolyMA-SCX (Ion Exchange) application for HPLC

CONDITIONS

column

Discovery BIO PolyMA-SCX, 5 cm x 4.6 mm, 5 μm particles (59601-U)

mobile phase

[A] 20 mM Bis-Tris HCl, pH 7.0; [B] 20 mM Bis-Tris HCl, 0.5M NaCl, pH 7.0; 24 to 69% B in 20 min. (linear)

flow rate

0.5 mL/min

column temp.

25 °C

detector

UV, 280 nm

injection

10 μg each variant

Description

Analysis Note

This Figure shows the separation of three cytochrome c variants on Discovery BIO PolyMA-SCX. Note the difference in elution order compared to the reversed-phase separation on Discovery BIO Wide Pore C18 shown in product G001581.

Legal Information

Discovery is a registered trademark of Merck KGaA, Darmstadt, Germany