Skip to Content
Merck

UHPLC/MS Analysis of Spice Cannabinoids on Titan C18

UHPLC/MS Analysis of Spice Cannabinoids on Titan™ C18 application for UHPLC

Materials

standard

Product No.
Description
Pricing

JWH-073 solution

100 μg/mL in acetonitrile, ampule of 1 mL, certified reference material, Cerilliant®

Spice Cannabinoid Mix 3 solution

100 μg/mL each component in acetonitrile, ampule of 1 mL, certified reference material, Cerilliant®

CONDITIONS

column

Titan C18, 10 cm x 2.1 mm, 1.9 μm particles (577124-U)

mobile phase

[A] 0.1% formic acid in 95:5, water:acetonitrile; [B] 0.1% formic acid in 5:95, water:acetonitrile;

gradient

25% B held for 1 min; to 45% B in 0.5 min; 45% B held for 4.5 min;   to 80% B in 0.5 min; 80% B held for 4 min

flow rate

0.5 mL/min

pressure

8756 psi (604 bar)

column temp.

35 °C

detector

MS-ESI+,100 - 1000 m/z scan, XIC

injection

2 μL

sample

100 ng/mL in 60:40, water:methanol

Description

Analysis Note

Synthetic cannabinoids (e.g. "Spice") are a type of designer drug that provides a cannabis-type high. New synthetic cannabinoids are continually being introduced as suppliers tweak the molecular structures. The ability to rapidly and reliably identify the continually changing population of these compounds is a significant analytical challenge facing forensic chemists. The rapid, efficient UHPLC separation a set of these compounds on a Titan C18 column is shown here. The highest grade LC-MS solvents were used to supply low background interference and low particulate contaminants for robust, trouble-free operation. Cerilliant CRMs provided reliable identification and quantification.

Legal Information

null