Accéder au contenu
Merck

HPLC Analysis of Withania Extract on Ascentis® Express Phenyl-Hexyl

HPLC Analysis of Withania Extract on Ascentis® Express Phenyl-Hexyl application for HPLC

Matériaux

Colonne analytique

Réf. du produit
Description
Tarif

Ascentis® Express Phenyl-Hexyl, 2.7 μm HPLC Column

2.7 μm particle size, L × I.D. 10 cm × 2.1 mm

Étalon

Réf. du produit
Description
Tarif

Withanoside IV

analytical standard

12-Deoxywithastramonolide

analytical standard

Withaferin A

analytical standard

CONDITIONS

column

Ascentis Express Phenyl-Hexyl, 10 cm x 2.1 mm I.D., 2.7 μm particles (53336-U)

mobile phase

[A] water; [B] acetonitrile

gradient

20 to 100% B in 10 min; held at 100% B for 0.5 min

flow rate

0.3 mL/min

pressure

160 bar (2320 psi)

column temp.

35 °C

detector

UV, 227 nm

injection

5 μL

sample

20 mg/mL in water:acetonitrile (50:50); sonicate 15 minutes; filter 0.45 μm;
dilute to water:acetonitrile (80:20)

Description

Remarque sur l'analyse

This application demonstrates the suitability of Ascentis Express Phenyl-Hexyl for the efficient separation of withania extract.

Informations légales

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany