Accéder au contenu
Merck

HPLC Analysis of Triacylglycerols in Soybean Oil on Discovery® HS C18 (Acetone-Acetonitrile Mobile Phase, ELSD Detection)

HPLC Analysis of Triacylglycerols in Soybean Oil on Discovery® HS C18 (Acetone-Acetonitrile Mobile Phase, ELSD Detection) application for HPLC

CONDITIONS

column

Discovery HS C18, 25 cm x 4.6 mm I.D., 5 μm particles (568523-U)

mobile phase

[A] acetone; [B] acetonitrile; (60:40, A:B)

flow rate

1.5 mL/min

pressure

914 psi (63 bar)

column temp.

30 °C

detector

ELSD

injection

5 μL

sample

5% soybean oil in acetone (w/v)

Description

Remarque sur l'analyse

Soybean oil was analyzed as 5% diluted sample in acetone (w/v)

Informations légales

Discovery is a registered trademark of Merck KGaA, Darmstadt, Germany