Accéder au contenu
Merck

HPLC Analysis of Beverage Additives in Diet Cola on Ascentis® Express RP-Amide

HPLC Analysis of Beverage Additives in Diet Cola on Ascentis® Express RP-Amide application for HPLC

Matériaux

Colonne analytique

Réf. du produit
Description
Tarif

Ascentis® Express RP-Amide, 2.7 μm HPLC Column

2.7 μm particle size, L × I.D. 10 cm × 3 mm

Étalon

Réf. du produit
Description
Tarif

Aspartame

analytical standard

Acide benzoique

analytical standard

Caféine

certified reference material, TraceCERT®, Manufactured by: Sigma-Aldrich Production GmbH, Switzerland

CONDITIONS

column

Ascentis Express RP-Amide, 10 cm x 3.0 mm I.D., 2.7 μm particles (53918-U)

mobile phase

[A] 20 mM ammonium acetate, pH 4.7 with acetic acid; [B] acetonitrile; (92:8, A:B)

flow rate

0.6 mL/min

column temp.

40 °C

detector

UV, 214 nm

injection

1 μL

sample

500 - 1000 μg/mL in 95:5, buffer: acetonitrile

Description

Remarque sur l'analyse

This application demonstrates the suitability of Ascentis Express RP-Amide for the efficient separation of common drink additives used in diet cola.

Informations légales

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany