- High-contrast grating resonators for label-free detection of disease biomarkers.
High-contrast grating resonators for label-free detection of disease biomarkers.
Scientific reports (2016-06-07)
Tianbo Sun, Shu Kan, Gerard Marriott, Connie Chang-Hasnain
PMID27265624
RÉSUMÉ
A label-free optical biosensor is described that employs a silicon-based high-contrast grating (HCG) resonator with a spectral linewidth of ~500 pm that is sensitive to ligand-induced changes in surface properties. The device is used to generate thermodynamic and kinetic data on surface-attached antibodies with their respective antigens. The device can detect serum cardiac troponin I, a biomarker of cardiac disease to 100 pg/ml within 4 mins, which is faster, and as sensitive as current enzyme-linked immuno-assays for cTnI.