Saltar al contenido
Merck

On-line continuous potentiometric measurement of potassium concentration in whole blood during open-heart surgery.

Clinical chemistry (1979-01-01)
H F Osswald, R Asper, W Dimai, W Simon
PMID761376
RESUMEN

We describe a flow-through system with an ion-selective electrode for measurement of blood potassium ion concentration, continuously and on-line off the extracorporeal blood circulation in an operating theater during human open-heart surgery. Comparison measurements were made with the SMA flame photometer (blood plasma) and an Orion SS 30 sodium/potassium analyzer (whole blood). The potassium concentration values obtained with the flow-through system agree well with the ones determined with the flame photometer. The time delay of the measurement with the flow-through system was relatively long (2 min) but delays of only 10--20 s seem feasible. Short time delays can deepen insight and simplify rational treatment under surgery conditions.

MATERIALES
Referencia del producto
Marca
Descripción del producto

Supelco
Potassium ionophore I, Selectophore, function tested
Número de referencia del producto (SKU)
Tamaño de envase
Disponibilidad
Precio
Cantidad