HPLC Analysis of an Energy Drink on Ascentis® Express HILIC (UV Detection)

HPLC Analysis of an Energy Drink on Ascentis® Express HILIC (UV Detection) application for HPLC

Materialien

Analytische Säule

Produkt-Nr.
Beschreibung
Preisangaben

Ascentis® Express HILIC HPLC Column

2.7 μm particle size, L × I.D. 10 cm × 3 mm

Standard

Produkt-Nr.
Beschreibung
Preisangaben

Koffein

Sigma Reference Standard, vial of 250 mg

Sorbinsäure

99.0-101.0% anhydrous basis

Pyridoxinhydrochlorid (B6)

analytical standard

Nicotinamid (Niacinamid)

analytical standard

Riboflavin (B2)

analytical standard

SS Benzoesäure

2000 μg/mL in dichloromethane, analytical standard

CONDITIONS

column

Ascentis Express HILlC, 10 cm x 3.0 mm I.D., 2.7 μm particles (53970-U)

mobile phase

[A] 100 mM ammonium acetate, pH 5.0 with acetic acid; [B] water; [C] acetonitrile; (09:01:90, A:B:C)

flow rate

0.6 mL/min

pressure

812 psi (56 bar)

column temp.

35 °C

detector

UV, 254 nm

injection

2 μL

sample

Dilute 1:9 in acetonitrile

Beschreibung

Hinweis zur Analyse

This application demonstrates the suitability of Ascentis Express HILIC for the efficient separation of common drink additives used in caffeinated energy drinks. The analysis was performed with UVand ELSD detectors in series, this chromatogram being from the UV.