GC Analysis of QA Test Mix on SLBĀ®-5ms, Three Different Phase Lots
Materialien
Analytische SƤule
Produkt-Nr.
Beschreibung
Preisangaben
CONDITIONS
column
SLB-5ms, 30 m x 0.25 mm I.D., 0.25 μm (28471-U)
oven
110 Ā°C (14 min), 15 Ā°C/min to 325 Ā°C (15 min)
inj. temp.
250 Ā°C
detector
FID, 360 Ā°C
carrier gas
helium, 30 cm/sec @ 110 Ā°C
injection
1.0 μL 100:1 split
liner
4 mm I.D. split, cup design
sample
Nonpolar Column Test Mix, each analyte at 500 μg/mL in methylene chloride
Beschreibung
Rechtliche Hinweise
SLB is a registered trademark of Merck KGaA, Darmstadt, Germany