G005921
GC Analysis of PCBs in Whole Cow Milk (Spiked) on SLB®-5ms (20 m x 0.18 mm I.D., 0.18 μm) after QuEChERS Cleanup using Supel™ QuE Z-Sep, Fast GC Analysis
suitable for GC, application for SPE
About This Item
Methode(n)
gas chromatography (GC): suitable
solid phase extraction (SPE): suitable
Testparameter
sample preparation: SPE (Solid Phase Extraction)
sample/matrix: 10 g whole cow′s milk
extraction tube: Supel™ QuE Non-buffered tube (4 g magnesium sulfate, 1 g sodium chloride) (55294-U)
clean-up tube: Supel™ QuE Z-Sep 150 mg, magnesium sulfate (150 mg, part no. 63135) (55299-U)
column: SLB-5ms, 20 m x 0.18 mm I.D., 0.18 μm (28564-U)
oven: 75 °C (1 min), 12 °C/min to 340 °C (10 min)
inj. temp.: 250 °C
detector: μ-ECD, 340 °C
carrier gas: hydrogen, 1.2 mL/min, constant
injection: 1 μL splitless (0.75 min)
liner: 4 mm I.D. FocusLiner™ with taper
Eignung
application for SPE
Anwendung(en)
food and beverages
Allgemeine Beschreibung
Hinweis zur Analyse
Extraction solvent: acetone:hexane (1:1), 20 mL
Clean-up volume: 1 mL
Rechtliche Hinweise
Choose from one of the most recent versions:
Analysenzertifikate (COA)
Sorry, we don't have COAs for this product available online at this time.
If you need assistance, please contact Kundensupport
Besitzen Sie dieses Produkt bereits?
In der Dokumentenbibliothek finden Sie die Dokumentation zu den Produkten, die Sie kürzlich erworben haben.
Unser Team von Wissenschaftlern verfügt über Erfahrung in allen Forschungsbereichen einschließlich Life Science, Materialwissenschaften, chemischer Synthese, Chromatographie, Analytik und vielen mehr..
Setzen Sie sich mit dem technischen Dienst in Verbindung.