G005711
LC/MS Analysis of Omeprazole and Metabolites in Plasma on Ascentis® Express RP Amide after Sample Prep using HybridSPE®-Phospholipid
application for HPLC, application for SPE
About This Item
Methode(n)
HPLC: suitable
solid phase extraction (SPE): suitable
Testparameter
sample preparation: SPE (Solid Phase Extraction)
sample/matrix: rat plasma spiked with omeprazole, phenacetin and metabolites at 2 μg/mL
SPE well plate: HybridSPE-Phospholipid, 96-well plate, 15 mg/0.8 mL (575656-U)
sample addition: 100 μL plasma to each well, followed by 300 mL of 1% formic acid in acetonitrile
elution: apply vacuum
eluate post-treatment: eluate was collected, evaporated, and reconstituted in mobile phase.
column: Ascentis Express RP-Amide, 5 cm x 2.1 mm I.D., 2.7 μm particles (53911-U)
mobile phase: [A] water, 10 mM ammonium formate, pH 3.4; [B] methanol, 10 mM ammonium formate, pH 3.4; (65:35, A:B)
flow rate: 0.3 mL/min
column temp.: 35 °C
detector: ESI(+), m/z = 50-400
injection: 1 μL
Eignung
application for HPLC
application for SPE
Anwendung(en)
clinical
Anwendung
Hinweis zur Analyse
Rechtliche Hinweise
Analyt
Analytische Säule
SPE-Röhrchen oder -Platte
Standard
Hier finden Sie alle aktuellen Versionen:
Analysenzertifikate (COA)
Leider sind derzeit keine COAs für dieses Produkt online verfügbar.
Wenn Sie Hilfe benötigen, wenden Sie sich bitte an Kundensupport
Besitzen Sie dieses Produkt bereits?
In der Dokumentenbibliothek finden Sie die Dokumentation zu den Produkten, die Sie kürzlich erworben haben.
Unser Team von Wissenschaftlern verfügt über Erfahrung in allen Forschungsbereichen einschließlich Life Science, Materialwissenschaften, chemischer Synthese, Chromatographie, Analytik und vielen mehr..
Setzen Sie sich mit dem technischen Dienst in Verbindung.