Passa al contenuto
Merck
Tutte le immagini(1)

Key Documents

G005602

HPLC Analysis of Triacylglycerols (TAG) of Borago Officinalis Oil on Ascentis® Express C18

application for HPLC

Autenticatiper visualizzare i prezzi riservati alla tua organizzazione & contrattuali


About This Item

tecniche

HPLC: suitable

Parametri analitici

column: Ascentis Express C18, 15 cm x 4.6 mm I.D., 2.7 μm particles (53829-U)
mobile phase: [A] acetonitrile; [B] isopropyl alcohol
gradient: 0% B held for 50 min; to 70% B in 5 min; to 0% B in 1 min
flow rate: 1 mL/min
detector: MS, APCI(+); mass spectral range, 250-1200 m/z; interval, 0.5 s; scan speed, 2143 amu/s; nebulizing gas (N2) flow, 4.0 L/min; interface temperature, 400 °C; (heat block, 230 °C; desolvation line (DL) temperature, 250 °C; DL voltage, 34 V; probe voltage, +4.5 kV; Qarray voltage, 1.0 V; RF voltage, 90 V; detection gain, 1.05 kV)
injection: 2 μL
sample: Borago Officinalis, 40 mg diluted in 10 mL of hexane

Compatibilità

application for HPLC

applicazioni

flavors and fragrances

Note legali

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany

Colonna analitica

N° Catalogo
Descrizione
Determinazione del prezzo

Scegli una delle versioni più recenti:

Certificati d'analisi (COA)

Lot/Batch Number

Ci dispiace, ma al momento non ci sono COA disponibili online per questo prodotto.

Se ti serve aiuto, non esitare a contattarci Servizio Clienti

Possiedi già questo prodotto?

I documenti relativi ai prodotti acquistati recentemente sono disponibili nell’Archivio dei documenti.

Visita l’Archivio dei documenti

Il team dei nostri ricercatori vanta grande esperienza in tutte le aree della ricerca quali Life Science, scienza dei materiali, sintesi chimica, cromatografia, discipline analitiche, ecc..

Contatta l'Assistenza Tecnica.