LC/MS Analysis of Illicit Bath Salts on Ascentis® Express HILIC, Set 2

LC/MS Analysis of Illicit Bath Salts on Ascentis® Express HILIC, Set 2 application for HPLC

CONDITIONS

column

Ascentis Express HILIC, 10 cm x 2.1 mm I.D., 2.7 μm particles (53939-U)

mobile phase

[A] 5 mM ammonium formate in acetonitrile: [B] water; (98:2, A:B)

flow rate

0.6 mL/min

pressure

267 psi (18.4 bar)

column temp.

35 °C

detector

MS, ESI(+) TOF, Summed Ion Chromatogram

injection

2 μL

sample

500 ng/mL in acetonitrile

Description

Remarque sur l'analyse

HILIC mode operation on an Ascentis Express column provided baseline resolution of these bath salts, including isobaric compounds. The highest quality LC-MS solvents were used to supply low background interference and low particulate contaminants for robust; trouble-free operation. Cerilliant CRMs provided reliable identification and quantification.