LC/MS/MS Analysis of Vitamin D Metabolites on Ascentis® Express F5

LC/MS/MS Analysis of Vitamin D Metabolites on Ascentis® Express F5 application for HPLC

Materials

analytical column

Product No.
Description
Pricing

Ascentis® Express F5, 2.7 μm HPLC Column

2.7 μm particle size, L × I.D. 15 cm × 2.1 mm

standard

Product No.
Description
Pricing

3-epi-25-Hydroxyvitamin D3

98% (CP)

3-epi-25-Hydroxyvitamin D2 solution

100 μg/mL in ethanol, 98% (CP)

25-Hydroxyvitamin D2 solution

50 μg/mL in ethanol, ampule of 1 mL, certified reference material, Cerilliant®

25-Hydroxyvitamin D3 solution

100 μg/mL in ethanol, ampule of 1 mL, certified reference material, Cerilliant®

1α,25-Dihydroxyvitamin D3 solution

5 μg/mL in ethanol, ampule of 1 mL, certified reference material, Cerilliant®

CONDITIONS

column

Ascentis Express  F5, 15 cm x 2.1 mm I.D., 2.7 μm particles (53571-U)

mobile phase

[A] 5 mM ammonium formate water; [B] methanol; (25:75, A:B)

flow rate

0.4 mL/min

pressure

5178 psi (357 bar)

column temp.

40 °C

detector

ESI(+), 100-1000 m/z

injection

2 μL, each compound 300 ng/mL in 40:60 (water:methanol)

Description

Analysis Note

The metabolic role and health implications of the various vitamin D isoforms are of current clinical interest. Therefore, it is important to have an analytical method that will resolve all of the known isoforms with necessary sensitivity and specificity.This application shows an Ascentis Express F5 provided rapid resolution of the isobaric vitamin D metabolites. MS detection provides the necessary secondary resolution. The highest grade LC-MS solvents were used to supply low background interference and low particulate contaminants for robust, trouble-free operation. Cerilliant CRMs provided reliable identification and quantification.